A modern, integrated service covering the design, manufacture, installation, and commissioning of analytical instruments for the petroleum and petrochemical industries.

+
  • 防爆离心机.png

Explosion-proof intelligent rapid determination instrument for crude oil water content (centrifugal method)

Category:

Product

Keywords:

Explosion-proof intelligent rapid determination instrument for crude oil water content (centrifugal method)

  • Commodity name: Explosion-proof intelligent rapid determination instrument for crude oil water content (centrifugal method)
  • Product Description
  • Technical parameters
  • Device maintenance and precautions
  • Explosion-proof certificate
    • Commodity name: Explosion-proof intelligent rapid determination instrument for crude oil water content (centrifugal method)
    • Commodity ID: 1354058113186680832

    This instrument uses a PLC control system and is explosion-proof. It mainly measures water and sediment in crude oil and is used in petroleum, chemical, scientific research, and oil product testing institutions.

  • Rated Voltage AC220V 50HZ

    Total Power 1.5KW

    Timing Range 1~ 99min

    Maximum Speed 5000r/min

    Sample Capacity: 0-60ml

    Accuracy: 0.05ml

    Speed Accuracy ± 25r/min

    Motor Power: 1KW

    Maximum Separation Factor 560/806

    Maximum Relative Centrifugal Force 3370xg

    Temperature Control Range Room Temperature: 65℃

    Temperature Control Accuracy ± 2

    Total Machine Noise 65dB A

  • 1. Do not move or open the door of the instrument while it is running.

    2. The instrument should be placed on a firm and level surface. All four rubber feet should be in contact with the surface and evenly distribute the weight to prevent vibration.

    3. The liquid in the centrifuge tubes should be balanced. Large differences in liquid levels will cause significant vibration during operation. Stop the machine and check the liquid levels to ensure they meet the requirements. Centrifuge tubes must be placed in even, symmetrical pairs.

    4. If a centrifuge tube breaks during operation, it will cause significant vibration. Stop the machine immediately and address the issue.

    5. The time interval between stopping and restarting the machine should not be less than 5 minutes to prevent compressor stalling and damage.

    6. After each centrifugation, the rotor must be removed. Leaving it on the shaft for extended periods may cause it to rust and become impossible to remove, resulting in the entire machine being scrapped.

    7. The power supply must be grounded.

    8. The service life of this machine's rotor is five years. The rotor should be replaced after this period.

    9. All rotors must not exceed their maximum speed.

     

  •  